Thermo Scientific Q Exactive MS System w/ Sogevac SV 40-65 Pump – Spares/Repair
This item was removed from a laboratory where it was surplus to requirement.
The system is in good cosmetic condition; however, an intermittent fault has been identified.
The previous owner advised that the unit developed an intermittent fault believed to be related to the electronics board controlling the instrument vacuum.
The vacuum system (turbo and roughing pumps) appears to function correctly based on instrument readbacks The system intermittently detects a loss of vacuum and shuts down to protect internal electronics (e.g. high voltage power supplies) The fault has occurred twice in the last month The instrument is able to re-establish vacuum without issue and has reportedly run for a week without recurrence
Due to this fault, the system is being sold for spares or repair.
Manufacturer
Thermo Scientific
Serial Number / VIN
ITEM-53802
Condition
Used - Good
Category
Cell Culture & Analysis
Seller
Typically replies within a few hours
Batch
#5088
Ask questions about this batch — pickup, condition, photos, or anything else.
Messages may be reviewed for trust & safety.