Thermo Scientific MaxQ 4000 Shaking Incubator – Spares/Repair. Removed from a laboratory environment where it was surplus to requirements. The unit powers on and is in good cosmetic condition with minor marks from use. Following inspection and basic fault-finding, power is confirmed as reaching the unit and the PSU is functioning correctly. However, power is not being passed through the control board to the PSU, indicating a likely fault with the power control board. The unit is therefore listed as Spares/Repair.